Title of article :
Chromatographic system for the simultaneous measurement of plasma 18-hydroxy-11-deoxycorticosterone and 18-hydroxycorticosterone by radioimmunoassay: reference data for neonates and infants and its application in aldosterone-synthase deficiency
Author/Authors :
Riepe، نويسنده , , F.G and Krone، نويسنده , , N and Peter، نويسنده , , M and Sippell، نويسنده , , W.G and Partsch، نويسنده , , C.-J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
A new chromatographic system for the steroid precursor separation and a sensitive radioimmunoassay system for the subsequent measurement of 18-hydroxy-11-deoxycorticosterone and 18-hydroxycorticosterone has been developed. 18-Hydroxy-11-deoxycorticosterone and 18-hydroxycorticosterone were extracted with methylene chloride and separated from cross-reacting steroids by Sephadex LH-20 column chromatography. Anti-18-hydroxy-11-deoxycorticosterone and anti-18-hydroxycorticosterone antibodies raised in rabbits were used. The lower detection limit of the assay is 0.03 nmol/l and 0.128 nmol/l for 18-hydroxy-11-deoxycorticosterone and 18-hydroxycorticosterone, respectively. Normal values for this assay in 128 healthy neonates and infants aged 0–5 months were established as a basis for the early hormonal diagnosis of aldosterone synthase deficiency types I and II. Its application for the diagnosis of aldosterone synthase deficiency is demonstrated in two patients with homozygous mutation/deletion in the encoding CYP11B2 gene.
Keywords :
18-Hydroxy-11-deoxycorticosterone , 18-Hydroxycorticosterone
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B