Title of article :
High-performance liquid chromatographic method with diode array detection to identify and quantify atypical antipsychotics and haloperidol in plasma after overdose
Author/Authors :
Titier، نويسنده , , K and Bouchet، نويسنده , , T and Pehourcq، نويسنده , , F and Moore، نويسنده , , N and Molimard، نويسنده , , M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
7
From page :
179
To page :
185
Abstract :
For toxicological purposes, an HPLC assay was developed for the simultaneous determination of haloperidol and atypical antipsychotics (risperidone, 9-hydroxyrisperidone, olanzapine, clozapine) in human plasma. After a double-step liquid–liquid extraction, compounds were separated on a C8 column eluted with a gradient of acetonitrile and phosphate buffer 50 mM pH 3.8. A sequential ultraviolet detection was used (260, 280 and 240 nm). Calibration curves were linear in the range 10–1000 ng/ml. The limits of quantification were 5 ng/ml for all drugs. Average accuracy at four concentrations ranged from 93 to 109%. Both inter- and intra-day variation coefficients were lower than 11% for all drugs. This simple and rapid method (run time<15 min) is currently used for poison management.
Keywords :
Haloperidol , 9-Hydroxyrisperidone , Clozapine , Risperidone , OLANZAPINE
Journal title :
Journal of Chromatography B
Serial Year :
2003
Journal title :
Journal of Chromatography B
Record number :
1455159
Link To Document :
بازگشت