Title of article :
Amperometric detection of perphenazine at a carbon fiber micro-disk bundle electrode by capillary zone electrophoresis
Author/Authors :
Liu، نويسنده , , Daxing and Jin، نويسنده , , Wenrui، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
A simple method for determination of perphenazine by capillary zone electrophoresis with amperometric detection is described. The optimum conditions of separation and detection are 1.50×10−3 mol/l Na2B4O7−1.0×10−3 mol/l NaOH (pH 9.9) for the buffer solution, 18 kV for the separation voltage, 5 kV and 5 s for the injection voltage and the injection time, and 0.80 V versus saturated calomel electrode for the detection potential, respectively. The limit of detection is 5.0×10−8 mol/l or 44 amol (S/N=3). The linear range of the calibration curve is 1.00×10−7 to 1.00×10−4 mol/l. The relative standard deviation is 1.5% for the migration time and 2.9% for the electrophoretic current at peak maximum. The method is applied to the determination of perphenazine in human urine.
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B