• Title of article

    Determination of tezosentan, a parenteral endothelin receptor antagonist, in human plasma by liquid chromatography–tandem mass spectrometry

  • Author/Authors

    van Giersbergen، نويسنده , , Paul L.M. and Wipfli، نويسنده , , Peter and Dingemanse، نويسنده , , Jasper، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    5
  • From page
    369
  • To page
    373
  • Abstract
    An analytical method was developed for the quantification of tezosentan in human plasma obtained in clinical studies. The method was linear in the range 1 to 512 ng/ml. After liquid–liquid extraction, the samples were analyzed by reversed-phase HPLC with tandem mass spectrometry. The limit of quantification was 1 ng/ml and the extraction recovery was at least 88.2%. Intra- and inter-assay coefficients of variation were below 10%. Stability tests revealed that tezosentan is stable under the different conditions tested.
  • Keywords
    Tezosentan
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography B
  • Record number

    1455627