• Title of article

    Automated large-volume sample stacking procedure to detect labeled peptides at picomolar concentration using capillary electrophoresis and laser-induced fluorescence detection

  • Author/Authors

    Siri، نويسنده , , Nathalie and Riolet، نويسنده , , Pierre and Bayle، نويسنده , , Christophe and Couderc، نويسنده , , François، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    151
  • To page
    157
  • Abstract
    We have developed an automated large-volume sample stacking (LVSS) procedure to detect fluorescein isothiocyanate-labeled peptides in the picomolar range. The injection duration is 10 min at 50 mbar to fill 62% of the capillary volume to the detection cell. The calculated limit of detection (S/N=3), filling 1% of the capillary volume, is 74 pM for bradykinin and 45 pM for l-enkephalin with samples diluted in water and analyzed in a 50 mM borate buffer, pH 9.2. With the automated LVSS system, the limits of detection are 7 pM for bradykinin, 3 pM for l-enkephalin and 2 pM for substance P. LVSS is shown to be quantitative from 500 to 10 pM.
  • Keywords
    Peptides , Fluorescein Isothiocyanate
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography B
  • Record number

    1455664