Title of article :
High-performance liquid chromatographic analysis of DA-7867, a new oxazolidinone, in human plasma and urine and in rat tissue homogenates
Author/Authors :
Bae، نويسنده , , Soo Kyung and Kim، نويسنده , , Eun Jung and Kwon، نويسنده , , Jong Won and Kim، نويسنده , , Won Bae and Lee، نويسنده , , Myung Gull Lee، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
An HPLC method was developed for the determination of a new oxazolidinone, DA-7867 (I), in human plasma and urine and in rat tissue homogenates. To 100 μl of biological sample, 300 μl acetonitrile and 50 μl methanol containing 10 μg/ml DA-7858 (the internal standard) were added. After vortex-mixing and centrifugation, the supernatant was evaporated under a gentle stream of nitrogen. The residue was reconstituted in 100 μl of the mobile phase and a 50-μl aliquot was injected directly onto the reversed-phase (C18) column. The mobile phase, 20 mM KH2PO4:acetonitrile (75:25, v/v) was run at a flow rate of 1.5 ml/min and the column effluent was monitored by a UV detector set at 300 nm. The retention times of I and DA-7858 were approximately 6.5 and 8.7 min, respectively. The detection limits of I in human plasma and urine and in rat tissue homogenates were 20, 20, and 50 ng/ml, respectively.
Keywords :
DA-7867 , oxazolidinones
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B