Title of article :
Determination of the antiangiogenesis agent 2-methoxyestradiol in human plasma by liquid chromatography with ultraviolet detection
Author/Authors :
Lakhani، نويسنده , , Nehal J and Sparreboom، نويسنده , , Alex and Dahut، نويسنده , , William L and Venitz، نويسنده , , Jürgen and Figg، نويسنده , , William D، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
5
From page :
289
To page :
293
Abstract :
A high-performance liquid chromatography (HPLC) assay was developed for the quantitative determination of 2-methoxyestradiol (2ME2) in human plasma. Sample pretreatment involved a solid-phase extraction of 1 ml aliquots of plasma with C18 micro-columns. Separation was achieved on a Novapak C18 column (300 mm×3.9 mm i.d.; 4 μm PS) at room temperature at an isocratic flow rate of 1 ml/min with 50% acetonitrile in water. Detection was performed at a UV wavelength of 205 nm. Calibration curves were linear in the concentration range of 1–50 ng/ml. The accuracy and precision values obtained from three different sets of quality controls analyzed in replicates of four on four separate occasions ranged from 90.7 to 105.2 and 3.17 to 8.27%, respectively.
Keywords :
2-Methoxyestradiol
Journal title :
Journal of Chromatography B
Serial Year :
2004
Journal title :
Journal of Chromatography B
Record number :
1456805
Link To Document :
بازگشت