Title of article :
Simple and sensitive fluorimetric liquid chromatography method for the determination of valproic acid in plasma
Author/Authors :
Lin، نويسنده , , Ming-Chun and Kou، نويسنده , , Hwang-Shang and Chen، نويسنده , , Cheng-Chung and Wu، نويسنده , , Shou-Mei and Wu، نويسنده , , Hsin-Lung، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A simple and sensitive liquid chromatographic method is described for the analysis of valproic acid in human plasma. The method is based on the derivatization of valproic acid extracted from acidified plasma with 2-(2-naphthoxy)ethyl 2-(piperidino)ethanesulfonate. The resulting derivative is highly responsive to a fluorimetric detector (excitation at 230 nm and emission at 350 nm), giving a low detection limit of 0.6 μM (S/N = 3, 10 μl injected). The relative standard deviations of the method for intra- and inter-day analyses (n = 5) are below 3.3 and 4.1%, respectively. Toluene was used for the extraction of valproic acid from plasma and the toluene extract obtained was subjected to subsequent derivatization without solvent replacement. The simple method was applied to the analysis of valproic acid in plasma of dosed patients using only small amount of sample (10–50 μl plasma).
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B