• Title of article

    High-performance liquid chromatography method for the quantification of pantoprazole in human plasma

  • Author/Authors

    Ramakrishna، نويسنده , , N.V.S and Vishwottam، نويسنده , , K.N and Wishu، نويسنده , , S. and Koteshwara، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    4
  • From page
    326
  • To page
    329
  • Abstract
    A sensitive and selective HPLC method with UV detection (290 nm) was developed and validated for quantitation of pantoprazole, proton-pump inhibitor, in human plasma. Following a single-step liquid–liquid extraction with methyl tert-butyl ether/diethyl ether (70/30, v/v), the analyte and internal standard (zonisamide) were separated using an isocratic mobile phase of 10 mM phosphate buffer (pH 6.0)/acetonitrile (61/39, v/v) on reverse phase Waters symmetry® C18 column. The lower limit of quantitation was 20 ng/mL, with a relative standard deviation of less than 4%. A linear range of 20–5000 ng/mL was established. This HPLC method was validated with between-batch and within-batch precision of 1.3–3.2% and 0.7–3.3%, respectively. The between-batch and within-batch bias was −0.5 to 8.2 % and −2.5 to 12.1%, respectively. This validated method is sensitive and repeatable enough to be used in pharmacokinetic studies.
  • Keywords
    Pantoprazole , HPLC , Quantification , Human plasma
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2005
  • Journal title
    Journal of Chromatography B
  • Record number

    1459148