Title of article :
Optimization of protein identification from digests as analyzed by capillary isoelectric focusing-mass spectrometry
Author/Authors :
Storms، نويسنده , , Henricus F. and van der Heijden، نويسنده , , Robert and Tjaden، نويسنده , , Ubbo R. and van der Greef، نويسنده , , Jan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Capillary isoelectric focusing (CIEF) is a high-resolution separation technique for the analysis of peptides and protein digests. When coupled to ion trap-mass spectrometry (CIEF-MS) the unique separation mechanism is combined with a highly efficient detection system. In an earlier report, we described aspects of separation and interfacing in connection to the analysis of a digest of set of standard proteins. Now, we report on different aspects of the process of protein identification. Sequest software parameters were optimized by using a standard protein digest. These settings were used for the analysis of periplasmic proteins from Escherichia coli. Since in CIEF peptides are focused according to their pI values, the mobilization time of a particular peptide is dependent on its pI value. Based on this relation, the identification of some peptides was facilitated. Furthermore, the Sequest settings that were used could be evaluated. In total, 159 proteins were identified in a single run.
Keywords :
Sequest , capillary isoelectric focusing , mass spectrometry , Data analysis , protein analysis
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B