• Title of article

    Ultra-sensitive determination of Formoterol in human serum by high performance liquid chromatography and electrospray tandem mass spectrometry

  • Author/Authors

    Mascher، نويسنده , , Daniel Gerhard and Zech، نويسنده , , Karl and Nave، نويسنده , , Rüdiger and Kubesch، نويسنده , , Klaus Michael and Mascher، نويسنده , , Hermann Josef، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    10
  • From page
    25
  • To page
    34
  • Abstract
    An analytical method was developed and validated to determine Formoterol in human serum in the range from 0.40 to 100.24 pg/mL by high performance liquid chromatography and tandem mass spectrometry (HPLC–MS/MS) due to the lack of efficient methods to determine very low levels of Formoterol in serum and plasma. Serum was diluted by water and mixed with the internal standard (d6-Formoterol). Formoterol and internal standard were extracted using a cation-exchange solid phase column (SCX-3). After eliminating endogenous serum constituents through washing steps with water and methanol, elution took place using methanol/ammonia. After evaporation of the elution liquid the residue was redissolved and analyzed by HPLC–MS/MS with electrospray ionisation (ESI) in positive mode. A gradient between 10 mM ammonium formate and acetonitrile was used. The inter-batch precision of the calibration standards ranged from 1.55% to 9.01%. The inter-batch accuracy of the calibration standards ranged from 93.37% to 107.30%. The lower limit of quantitation (LLOQ, 0.40 pg/mL) had a precision of 19.67% and an accuracy of 96.78%. Comparable results were obtained for quality control samples. Stability in human serum was given over three freeze/thaw cycles and 2 h at room temperature. Formoterol in human serum was stable for at least 6 months below −20 °C. This method has been used widely for quantifying Formoterol after inhalation of 9–36 μg of the drug by volunteers. A cross validation with human plasma versus serum was performed after this method was successfully validated in human serum.
  • Keywords
    Formoterol , Ultra sensitive method , Tandem mass spectrometry , Human serum , Solid phase extraction , Liquid chromatography
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography B
  • Record number