Title of article
Sensitive column-switching high-performance liquid chromatography method for determination of propiverine in human plasma
Author/Authors
Ban، نويسنده , , Eunmi and Maeng، نويسنده , , Jeong-Eun and Woo، نويسنده , , Jong Soo and Kim، نويسنده , , Chong-Kook Kim، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
6
From page
230
To page
235
Abstract
A sensitive column-switching high-performance liquid chromatography (HPLC) with ultraviolet (UV) detection was developed for the determination of propiverine in human plasma. Propiverine and internal standard, oxybutynin, were extracted from human plasma that had been made basic with 5N sodium hydroxide into methyl tert-butyl ether. The extracted plasma sample was injected onto the HPLC system consisting of a pretreatment column, a concentrating column, and an analytical column, which were connected with a six-port switching valve. The assay was linear in concentration ranges of 2–200 ng/ml for propiverine in human plasma. This method showed excellent sensitivity (a limit of detection of 0.5 ng/ml), good precision and accuracy. This method is suitable for bioequivalence studies following single dose in healthy volunteers.
Keywords
Propiverine , Column switching , High-Performance Liquid Chromatography (HPLC)
Journal title
Journal of Chromatography B
Serial Year
2006
Journal title
Journal of Chromatography B
Record number
1462729
Link To Document