• Title of article

    Liquid chromatography–tandem mass spectrometry assay of reduced and oxidized glutathione and main precursors in mice liver

  • Author/Authors

    Bouligand، نويسنده , , Jérôme and Deroussent، نويسنده , , Alain and Paci، نويسنده , , Angelo and Morizet، نويسنده , , Jackie and Vassal، نويسنده , , Gilles، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    8
  • From page
    67
  • To page
    74
  • Abstract
    A liquid chromatography/tandem mass spectrometry assay of glutathione (GSH), glutathione disulfide (GSSG) and of precursors (γ-glutamyl-cysteine, cysteinyl-glycine, cysteine, cystine, homocysteine and homocystine) was developed to study glutathione synthesis in mice liver. After iodoacetic acid derivatization, the analytes were analyzed using reversed-phase gradient HPLC and detected using multiple reaction monitoring. Linear calibrations were performed over the concentrations range of 100–10,000 ng/mL for the thiol-containing precursors and extended up to 100,000 ng/mL for GSH and GSSG. The method was validated for each compound with inter-day accuracy below 11.9% and with precision below 15%. The method showed low limits of quantitation of 100 ng/mL for each thiol-containing compound and GSSG and of 200 ng/mL for other disulfides.
  • Keywords
    Disulfide glutathione , HPLC–MS/MS , Liver , Glutathione precursors , mice , glutathione
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography B
  • Record number

    1462797