Title of article :
Liquid chromatography–electrospray ionization mass spectrometry for direct identification and quantification of iophenoxic acid in serum
Author/Authors :
Wiles، نويسنده , , Melinda C. and Campbell، نويسنده , , Tyler A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
14
From page :
144
To page :
157
Abstract :
A liquid chromatographic–electrospray ionization mass spectrometric technique was developed for direct quantitation of iophenoxic acid (IA) in serum. IA was spiked into canine, feline, bovine, equine, and porcine sera, extracted, and quantified using negative ion monitoring following chromatographic separation on a Luna C18(2) 3 μm (100 mm × 2.1 mm) reversed-phase column. The limit of detection was 25 ng/mL and the limit of quantification was 50 ng/mL. Inter- and intra-assay accuracy (86–113% and 87–115%, respectively) and precision (1.8–7.7%) were calculated. Analysis of serum collected from feral pigs, raccoons, and opossums following ingestion of IA-marked baits confirmed the appropriateness of this method for bait acceptance studies.
Keywords :
LC–MS , Negative ion monitoring , electrospray ionization , Iophenoxic acid , Bait acceptance , serum
Journal title :
Journal of Chromatography B
Serial Year :
2006
Journal title :
Journal of Chromatography B
Record number :
1462825
Link To Document :
بازگشت