• Title of article

    Quantification of uric acid, xanthine and hypoxanthine in human serum by HPLC for pharmacodynamic studies

  • Author/Authors

    Cooper، نويسنده , Paul W , Nancy and Khosravan، نويسنده , , Reza and Erdmann، نويسنده , , Carol and Fiene، نويسنده , , John and Lee، نويسنده , , Jean W.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    10
  • From page
    1
  • To page
    10
  • Abstract
    A simple HPLC method was developed and validated for the determination of uric acid (UA), xanthine (X) and hypoxanthine (HX) concentrations in human serum to support pharmacodynamic (PD) studies of a novel xanthine oxidase inhibitor during its clinical development. Serum proteins were removed by ultrafiltration. The hydrophilic analytes and the I.S. were eluted by 100% aqueous phosphate buffer mobile phase. The hydrophobic matrix components (late peaks) were eluted with a step gradient of a higher organic mobile phase. Validation on linearity, sensitivity, precision, accuracy, stability, and robustness of the method for PD biomarkers (UA, X, and HX) was carried out in a similar manner to that for pharmacokinetic (PK) data where applicable. Issues of selectivity for endogenous biomarker analytes and individual concentration variations were addressed during method validation. Standards were prepared in analyte-free phosphate buffer. Quality control samples were prepared in control serum from individuals not dosed with the xanthine oxidase inhibitor. The method was simple and robust with good accuracy and precision for the measurement of serum UA, X, and HX concentrations.
  • Keywords
    uric acid , Xanthine oxidase inhibitor , HPLC , Pharmacodynamic biomarkers , Human serum
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography B
  • Record number

    1463158