Title of article :
On-line biological sample cleanup for electrospray mass spectrometry using sol–gel columns
Author/Authors :
Johannesson، نويسنده , , Nina and Pearce، نويسنده , , Evan and Dulay، نويسنده , , Maria and Zare، نويسنده , , Richard N. and Bergquist، نويسنده , , Jonas and Markides، نويسنده , , Karin E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
5
From page :
70
To page :
74
Abstract :
Using a slight overpressure, a urine sample is loaded onto a monolithic photopolymerized sol–gel column that has been derivatized with hydrophobic carbon chains and then the complex urine matrix is washed with aqueous solution. A buffer containing organic solvent is used to elute the adsorbed peptides by an applied voltage and the sample is then introduced into a mass spectrometer by sheath flow electrospray. The importance of desalting this type of sample is demonstrated by an experiment that shows that the signal intensity of a test solution with neurotensin, sprayed directly into the mass spectrometer, decreased from 4.5 × 104 cps to no detectible signal when just 10% urine is added to the sample solution. We suggest that this procedure may find general application for desalting biological samples prior to mass spectrometric analysis.
Keywords :
mass spectrometry , Preconcentration , ON-LINE , electrophoresis , Biological sample , Octadecyltrimethoxysilane , desalting , Sol–gel