Title of article :
A simple and rapid determination of biapenem in plasma by high-performance liquid chromatography
Author/Authors :
Ikeda، نويسنده , , Kayo and Ikawa، نويسنده , , Kazuro and Ikeda، نويسنده , , Aki and Nishikawa، نويسنده , , Yoshimi and Morikawa، نويسنده , , Norifumi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
A simple, rapid and precise HPLC method using ultrafiltration to remove plasma protein was developed to determine biapenem concentrations in human plasma. Plasma was separated by centrifugation at 4 °C from blood collected in heparinized vacuum tubes, and biapenem was stabilized by immediate mixing the plasma with 1 M 3-morpholinopropanesulfonic acid (MOPS) buffer (pH 7.0) (1:1). Biapenem was detected by ultraviolet absorbance at 300 nm with no interfering plasma peak. The calibration curve of biapenem in human plasma was linear from 0.04 to 50 μg/mL. The limit of detection was 0.01 μg/mL, which was more than 40-fold lower than that of conventional plasma protein precipitation using ammonium sulfate. The assay has been clinically applied to pharmacokinetic studies in patients.
Keywords :
Biapenem , Ultrafiltration
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B