Title of article
Novel generic UPLC/MS/MS method for high throughput analysis applied to permeability assessment in early Drug Discovery
Author/Authors
Mensch، نويسنده , , Jurgen and Noppe، نويسنده , , Mark and Adriaensen، نويسنده , , Jef and Melis، نويسنده , , Anouche and Mackie، نويسنده , , Claire and Augustijns، نويسنده , , Patrick and Brewster، نويسنده , , Marcus E، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
6
From page
182
To page
187
Abstract
A novel generic ultra performance liquid chromatography-tandem mass spectrometric (UPLC/MS/MS) method for the high throughput quantification of samples generated during permeability assessment (PAMPA) has been developed and validated. The novel UPLC/MS/MS methodology consists of two stages. Firstly, running a 1.5 min isocratic method, compound-specific multiple reaction monitoring (MRM) methods were automatically prepared. In a second stage, samples were analyzed by a 1.5 min generic gradient UPLC method on a BEH C18 column (50 mm × 2.1 mm). Compounds were detected with a Waters Micromass Quattro Premier mass spectrometer operating in positive electrospray ionization using the compound-specific MRM methods. The linearity for the validation compounds (caffeine, propranolol, ampicillin, atenolol, griseofulvin and carbamazepine) typically ranges from 3.05 nM to 12,500 nM and the limits of detection for all generically developed methods are in the range between 0.61 nM and 12 nM in an aqueous buffer. The novel generic methodology was successfully introduced within early Drug Discovery and resulted in a four-fold increase of throughput as well as a significant increase in sensitivity compared to other in-house generic LC/MS methods.
Keywords
Permeability , PAMPA , high throughput , UPLC , UPLC/MS/MS , MRM , Quanoptimize
Journal title
Journal of Chromatography B
Serial Year
2007
Journal title
Journal of Chromatography B
Record number
1463946
Link To Document