Title of article :
The quantification of erlotinib (OSI-774) and OSI-420 in human plasma by liquid chromatography–tandem mass spectrometry
Author/Authors :
Masters، نويسنده , , Andrea R. and Sweeney، نويسنده , , Christopher J. H. Jones، نويسنده , , David R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
5
From page :
379
To page :
383
Abstract :
An accurate and precise method was developed using HPLC-MS/MS to quantify erlotinib (OSI-774) and its O-desmethyl metabolite, OSI-420, in plasma. The advantages of this method include the use of a small sample volume, liquid–liquid extraction with high extraction efficiency and short chromatographic run times. The analytes were extracted from 100 μL plasma volume using hexane:ethyl acetate after midazolam was added to the sample for internal standardization. The compounds were separated on a Phenomenex C-18 Luna analytical column with acetonitrile:5 mM ammonium acetate as the mobile phase. All compounds were monitored by tandem mass spectrometry with electrospray positive ionization. The intra-day accuracy and precision (% coefficient of variation, % CV) estimates for erlotinib at 10 ng/mL were 90% and 9%, respectively. The intra-day accuracy and precision estimates for OSI-420 at 5 ng/mL were 80% and 4%, respectively. This method was used to quantify erlotinib and OSI-420 in plasma of patients (n = 21) administered 150 mg erlotinib per day for non-small cell lung cancer.
Keywords :
Erlotinib , OSI-774 , HPLC-MS/MS
Journal title :
Journal of Chromatography B
Serial Year :
2007
Journal title :
Journal of Chromatography B
Record number :
1464094
Link To Document :
بازگشت