Title of article
Matrix-comprehensive in-house validation and robustness check of a confirmatory method for the determination of four nitrofuran metabolites in poultry muscle and shrimp by LC–MS/MS
Author/Authors
Bock، نويسنده , , Claudia and Gowik، نويسنده , , Petra and Stachel، نويسنده , , Carolin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
12
From page
178
To page
189
Abstract
An already well-described method for the determination of nitrofuran metabolites 3-amino-5-methyl-morpholino-2-oxazolidinone (AMOZ), 3-amino-2-oxazolidinone (AOZ), semicarbazide (SEM) and 1-aminohydantoin (AHD) was adapted to the needs of our laboratory and checked for its robustness regarding sample conditions and the processing step. Using the same data, the method was validated and the measurement uncertainty was estimated. All criteria and requirements of Commission Decision 2002/657/EC were fulfilled. The CCα determined lies between 0.1 and 0.7 μg/kg, the CCβ lies between 0.1 and 0.9 μg/kg, the measurement uncertainty was estimated as being between 7 and 17% taking into account matrix, time and sample preparation influences.
Keywords
Validation , Shrimps , Muscle , LC–MS/MS , Residues , measurement uncertainty , Matrix-comprehensive in-house validation , metabolite , Robustness , Poultry , Nitrofurans