• Title of article

    Liquid chromatography tandem mass spectrometry method for simultaneous determination of metoprolol tartrate and ramipril in human plasma

  • Author/Authors

    Gowda، نويسنده , , K. Veeran and Mandal، نويسنده , , Uttam and Senthamil Selvan، نويسنده , , P. and Sam Solomon، نويسنده , , W.D. and Ghosh، نويسنده , , Animesh and Sarkar، نويسنده , , Amlan Kanti and Agarwal، نويسنده , , Sangita and Nageswar Rao، نويسنده , , T. and Pal، نويسنده , , Tapan Kumar، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    9
  • From page
    13
  • To page
    21
  • Abstract
    A simple, rapid, sensitive and specific liquid chromatography-tandem mass spectrometry method was developed and validated for quantification of metoprolol tartrate (MT) and ramipril, in human plasma. Both the drugs were extracted by liquid–liquid extraction with diethyl ether–dichloromethane (70:30, v/v). The chromatographic separation was performed on a reversed-phase C8 column with a mobile phase of 10 mM ammonium formate–methanol (3:97, v/v). The protonated analyte was quantitated in positive ionization by multiple reaction monitoring with a mass spectrometer. The method was validated over the concentration range of 5–500 ng/ml for metoprolol and ramipril in human plasma. The precursor to product ion transitions of m/z 268.0–103.10 and m/z 417.20–117.20 were used to measure metoprolol and ramipril, respectively.
  • Keywords
    Ramipril , Metoprolol tartrate , Validation , LC–MS/MS
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2007
  • Journal title
    Journal of Chromatography B
  • Record number