Title of article :
Validation and application of a 96-well format solid-phase extraction and liquid chromatography–tandem mass spectrometry method for the quantitation of digoxin in human plasma
Author/Authors :
Hashimoto، نويسنده , , Yoshitaka and Shibakawa، نويسنده , , Kimio and Nakade، نويسنده , , Susumu and Miyata، نويسنده , , Yasuyuki، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
To evaluate the pharmacokinetics of digoxin in humans, a sensitive and specific LC/MS/MS method was developed and validated for the determination of digoxin concentrations in human plasma. The method was shown to be more sensitive, specific, accurate, and reproducible than common techniques such as RIA. For detection, a LC/MS/MS system with electro spray ionization tandem mass spectrometry in the positive ion-multiple reaction-monitoring (MRM) mode was used to monitor precursor to product ions of m/z 798.5–51.5 for digoxin and m/z 782.5–35.5 for the internal standard, digitoxin. The method was validated over a concentration range of 0.02–5 ng/mL and was found to have acceptable accuracy, precision, linearity, and selectivity. The mean extraction recovery from spiked plasma samples was above 80%. Imidafenacin, coadministered in a drug–drug interaction study, had no detectable influence on the determination of digoxin in human plasma. The novel method was applied to a drug–drug interaction study of digoxin and imidafenacin and the characterization of steady-state pharmacokinetics of digoxin in humans after oral administration at a dose of 0.25 mg on days 1 and 2 followed by 0.125 mg daily doses on days 3 through 8.
Keywords :
LC/MS/MS , 96-Well SPE , Imidafenacin , Digoxin , Human plasma
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B