Title of article :
Validated high performance liquid chromatography–UV detection method for the determination of daptomycin in human plasma
Author/Authors :
Jens Martens-Lobenhoffer، نويسنده , , Jens and Kielstein، نويسنده , , Jan T. and Oye، نويسنده , , Catrin and Bode-Bِger، نويسنده , , Stefanie M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Daptomycin is the first approved member of the new class cyclic lipopeptide antibiotic drugs, effective against a broad spectrum of Gram-positive bacteria. Here we present an HPLC method with UV detection capable to obtain pharmacokinetic data of daptomycin in human plasma, exemplarily shown in a critically ill patient with acute renal failure undergoing extended daily dialysis. Sample preparation consists only of protein precipitation with methanol. Chromatographic separation was achieved on a Zorbax Eclipse XDB-C8 column and daptomycin was detected at 224 nm. The calibration function was linear over the range from 3.5 to 350 μg/ml. The relative standard deviations were < 2% in the intra-day and < 6% in the inter-day measurements. The accuracy was always better than 7%. Daptomycin was stable in aqueous solutions for 2 months frozen at–20 °C. However, in plasma frozen at −20 °C a loss of 25% in 1 month was observed.
Keywords :
Daptomycin , UV , stability , HPLC
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B