Title of article
Detection and quantification of fatty acid ethyl esters in meconium by headspace-solid-phase microextraction and gas chromatography–mass spectrometry
Author/Authors
Hutson، نويسنده , , J.R. and Aleksa، نويسنده , , K. and Pragst، نويسنده , , F. and Koren، نويسنده , , G.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
5
From page
8
To page
12
Abstract
Meconium fatty acid ethyl esters (FAEEs) are currently used as biomarkers to detect heavy prenatal alcohol exposure. We introduce a novel technique to quantify FAEEs in meconium using headspace-solid-phase microextraction (HS-SPME) coupled with gas chromatography–mass spectrometry (GC–MS). This method improves on previous approaches by decreasing sample preparation time, eliminating the need for organic solvents, and reducing the required sample size. Using 50 mg of meconium, the detection limits of FAEEs ranged from 0.05 to 0.16 nmol/g and had good reproducibility making it ideal for routine analysis of clinical samples.
Keywords
Meconium , Fatty acid ethyl esters , Gas chromatography , Headspace-solid-phase microextraction , Biomarker
Journal title
Journal of Chromatography B
Serial Year
2009
Journal title
Journal of Chromatography B
Record number
1466699
Link To Document