• Title of article

    Isocratic rapid liquid chromatographic method for simultaneous determination of carotenoids, retinol, and tocopherols in human serum

  • Author/Authors

    Thibeault، نويسنده , , Denis and Su، نويسنده , , Haixiang and MacNamara، نويسنده , , Elizabeth and Schipper، نويسنده , , Hyman M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    7
  • From page
    1077
  • To page
    1083
  • Abstract
    An improved isocratic and rapid HPLC method was developed for the measurement of carotenoids, retinol and tocopherols in human serum. Vitamins were extracted with hexane. Mobile phase consisted of a mixture acetonitrile:methylene chloride:methanol with 20 mM ammonium acetate. This method used a small bead size (3 μm) Spherisorb ODS2 column with titane frits. Diode array and fluorescence detectors were used respectively for the detection of carotenoids and retinol/tocopherols. Chromatographic separation was complete in 13 min for β-cryptoxanthin, cis–trans-lycopene, α-carotene, β-carotene, cis-β-carotene, retinol, δ-tocopherol, γ-tocopherol and α-tocopherol. Echinenone and tocol were employed as internal standards for diode array and fluorescence detection. Accuracy was validated using standard reference material (SRM) 968C. Intra-assay and inter-assay precision were respectively 0.2–7.3% and 3.6–12.6%. Sensitivity was verified using the ICH recommendations and the limit of detection (LOD) obtained was sufficient for routine clinical application.
  • Keywords
    Tocopherols , Diode array detector , HPLC , Fluorescence detector , fat-soluble vitamins , Retinol , carotenoids
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography B
  • Record number

    1467064