• Title of article

    Development and validation of a sensitive and rapid non-aqueous LC–ESI-MS/MS method for measurement of diosgenin in the plasma of normal and hyperlipidemic rats: A comparative study

  • Author/Authors

    Xu، نويسنده , , Lina and Liu، نويسنده , , Yuetao and Wang، نويسنده , , Tie and Qi، نويسنده , , Yan and Han، نويسنده , , Xu and Xu، نويسنده , , Youwei and Peng، نويسنده , , Jinyong and Tang، نويسنده , , Xinqiang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    7
  • From page
    1530
  • To page
    1536
  • Abstract
    A sensitive and specific electrospray ionization liquid chromatography–tandem mass spectrometry method was developed to detect diosgenin in the plasma of normal and hyperlipidemic rats. Diosgenin was extracted with n-hexane–ethyl acetate (9:1, v/v) using sarsasapogenin as an internal standard. With multiple reaction monitoring modes, linear calibration curves were obtained in the range 10–1500 ng/mL (r ≥ 0.9979) and the limit of quantification was 10 ng/mL. Intra- and inter-assay variabilities were within 7.74%, and accuracies were between −5.33% and 1.50%. The assay was successfully applied to study pharmacokinetics in rats after oral administration of diosgenin. Significantly different pharmacokinetics between normal and hyperlipidemic rats were observed, which would be beneficial for the clinical use of diosgenin.
  • Keywords
    Pharmacokinetics , Diosgenin , Hyperlipidemic rat , Sarsasapogenin , LC–MS/MS
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography B
  • Record number