• Title of article

    Development of a high performance anion exchange chromatography analysis for mapping of oligosaccharides

  • Author/Authors

    Grey، نويسنده , , Carl and Edebrink، نويسنده , , Per and Krook، نويسنده , , Maria and Jacobsson، نويسنده , , Sven P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    1827
  • To page
    1832
  • Abstract
    In the present study a HPAEC-PAD method is described that was developed for monitoring the consistency of N-glycosylation during the production and purification of recombinant proteins and monoclonal antibodies. The method successfully separated 18 neutral and sialylated oligosaccharides. Results obtained were compared with MALDI-TOF MS and it was shown that both methods gave similar results. In addition, a method validation was performed showing that the HPAEC-PAD analysis was well suited for the mapping and characterization of oligosaccharides. The method was found to be robust and additionally the precision was significantly better compared to the MALDI-TOF MS method.
  • Keywords
    N-glycosylation , Monoclonal antibodies , MALDI-TOF MS , glycoprotein , oligosaccharides , HPAEC-PAD
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography B
  • Record number

    1467303