Title of article :
A sensitive LC–MS/MS method for quantification of a nucleoside analog in plasma: Application to in vivo rat pharmacokinetic studies
Author/Authors :
Olivier Heudi، نويسنده , , Olivier and Barteau، نويسنده , , Samuel and Picard، نويسنده , , Franck and Kretz، نويسنده , , Olivier، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
A LC–MS/MS method was developed and validated for determination of nucleoside analog (NA) in rat plasma. The method run time was 6 min and the limit of quantification (LOQ) was estimated at 100 pg/mL. The extraction procedure consisted on plasma samples protein precipitation with an acetonitrile solution which contained the stable isotope labeled internal standard (IS). Chromatography was performed on a newly developed C16 column (150 mm × 4.6 mm, 5 μm) in order to avoid the use ion pair salts. The samples were eluted at 0.8 mL/min with a gradient of mobile phase made of water and acetonitrile both acidified with 0.5% acetic acid and 0.025% trifluoroacetic acid (TFA). A tandem mass spectrometer was used as a detector for quantitative analysis. Intra-run and inter-run precision and accuracy within ±15% were achieved during a 3-run validation for quality control samples at four concentration levels in rat plasma, over a concentration ranging between 0.1 and 1000 ng/mL. The data indicate that our LC–MS/MS assay is an effective method for the pharmacokinetics study of NA in rat plasma.
Keywords :
nucleoside analogs , LC–MS/MS , Polar compounds , TFA , Validation and pharmacokinetic
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B