Title of article :
A validated assay for the quantitative analysis of vatalanib in human EDTA plasma by liquid chromatography coupled with electrospray ionization tandem mass spectrometry
Author/Authors :
Lankheet، نويسنده , , A.G. and Hillebrand، نويسنده , , M.J.X. and Langenberg، نويسنده , , M.H.G. and Rosing، نويسنده , , Jos H. and Huitema، نويسنده , , A.D.R. and Voest، نويسنده , , E.E. and Schellens، نويسنده , , J.H.M. and Beijnen، نويسنده , , J.H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
6
From page :
3625
To page :
3630
Abstract :
A sensitive and accurate method for the determination of vatalanib in human EDTA plasma was developed using high-performance liquid chromatography and detection with tandem mass spectrometry. Stable isotopically labeled imatinib was used as internal standard. Plasma proteins were precipitated and an aliquot of the supernatant was directly injected onto a Phenomenex Gemini C18 analytical column (50 mm × 2.0 mm ID, 5.0 μm particle size) and then compounds were eluted with a linear gradient. The outlet of the column was connected to a Sciex API 365 triple quadrupole mass spectrometer and ions were detected in positive multiple reaction monitoring mode. The lower limit of quantification was 10 ng/mL (S/N ≈ 10, CV ≤ 8.4%). This method was validated over a linear range from 10 to 2500 ng/mL, and results from the validation study demonstrated a good intra- and inter-assay accuracy (inaccuracy ≤9.57%) and precision (CV ≤ 8.81%). This method has been used to determine plasma vatalanib concentrations in patients with advanced solid tumor, enrolled in a phase I pharmacokinetic trial with the drug.
Keywords :
Vatalanib , mass spectrometry , Liquid chromatography
Journal title :
Journal of Chromatography B
Serial Year :
2009
Journal title :
Journal of Chromatography B
Record number :
1467799
Link To Document :
بازگشت