Validation and application of a liquid chromatography–tandem mass spectrometric method for quantification of the drug transport probe fexofenadine in human plasma using 96-well filter plates
Author/Authors :
Stanton، نويسنده , , Melonie L. and Joy، نويسنده , , Melanie S. and Frye، نويسنده , , Reginald F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
A rapid method to determine fexofenadine concentrations in human plasma using protein precipitation in 96-well plates and liquid chromatography–tandem mass spectrometry was validated. Plasma proteins were precipitated with acetonitrile containing the internal standard fexofenadine-d6, mixed briefly, and then filtered into a collection plate. The resulting filtrate was diluted and injected onto a Phenomenex Gemini C18 (50 mm × 2.0 mm, 5 μm) analytical column. The mobile phase consisted of 0.1% formic acid, 5 mM ammonium acetate in deionized water and methanol (35:65, v/v). The flow rate was 0.2 ml/min and the total run time was 2 min. Detection of the analytes was achieved using positive ion electrospray ionization and high resolution multiple reaction monitoring mode (H-SRM). The linear standard curve ranged from 1 to 500 ng/ml and the precision and accuracy (intra- and inter-run) were within 4.3% and 8.0%, respectively. The method has been applied successfully to determine fexofenadine concentrations in human plasma samples obtained from subjects administered a single oral dose of fexofenadine. The method is rapid, sensitive, selective and directly applicable to human pharmacokinetic studies involving fexofenadine.
Keywords :
fexofenadine , Fexofenadine-d6 , Protein precipitation , Human plasma , LC–MS
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B