• Title of article

    A sensitive and selective liquid chromatography/tandem mass spectrometry method for determination of MLN8237 in human plasma

  • Author/Authors

    Lipsitz، نويسنده , , Emily and Moorthy، نويسنده , , Ganesh and Mosse، نويسنده , , Yael and Fox، نويسنده , , Elizabeth and Adamson، نويسنده , , Peter C.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    5
  • From page
    2369
  • To page
    2373
  • Abstract
    We describe a selective and a highly sensitive high-performance liquid chromatography–electron spray ionization-collision induced dissociation-tandem mass spectrometry (HPLC–ESI-CID-MS/MS) assay for the Aurora A kinase inhibitor MLN8237 in human plasma. The intra-day precision based on the standard deviation of replicates of quality control samples ranged from 0.2 to 4% and with accuracy ranging from 96 to 102%. The inter-day precision ranged from 0.5 to 7% and the accuracy ranged from 93 to 105%. Stability studies showed that MLN8237 was stable both during the expected conditions for sample preparation and storage. The lower limit of quantification for MLN8237 was 5 ng/mL. The analytical method showed excellent sensitivity, precision, and accuracy. This method is robust and is being successfully employed in a Childrenʹs Oncology Group Phase 1 Consortium study of MLN8237 in children with cancer.
  • Keywords
    MLN8237 , PLASMA , CANCER , LC–ESI-CID-MS/MS , Pediatric
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography B
  • Record number