Title of article :
Evaluation of the performance of protein separation in figure-8 centrifugal counter-current chromatography
Author/Authors :
Yang، نويسنده , , Yi-jie GU، نويسنده , , Dongyu and Aisa، نويسنده , , Haji Akber and Ito، نويسنده , , Yoichiro، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
4
From page :
3802
To page :
3805
Abstract :
The performance of protein separation using the figure-8 column configuration in centrifugal counter-current chromatography was investigated under various flow rates and revolution speeds. The separation was performed with a two-phase solvent system composed of polyethylene glycol 1000/potassium phosphate each at 12.5% (w/w) in water and with lysozyme and myoglobin as test samples. In order to improve tracing of the elution curve, a hollow fiber membrane dialyzer was inserted at the inlet of the UV detector. The results showed that the retention of stationary phase (Sf) and resolution (Rs) increased with decreased flow rate and increased revolution speed. The highest Rs of approximately 1 was obtained at a flow rate of 0.01 mL/min under a revolution speed of 1200 rpm with a 3.4 mL capacity column.
Keywords :
Centrifugal counter-current chromatography , Figure-8 column , Protein Separation , Polymer phase system , Lysozyme , myoglobin , Hollow fiber membrane dialyzer
Journal title :
Journal of Chromatography B
Serial Year :
2011
Journal title :
Journal of Chromatography B
Record number :
1469235