Title of article :
Liquid chromatography–tandem mass spectrometric assay for the JAK2 inhibitor CYT387 in plasma
Author/Authors :
Sparidans، نويسنده , , Rolf W. and Durmus، نويسنده , , Selvi and Xu، نويسنده , , Ning and Schinkel، نويسنده , , Alfred H. and Schellens، نويسنده , , Jan H.M and Beijnen، نويسنده , , Jos H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
4
From page :
174
To page :
177
Abstract :
A quantitative bioanalytical liquid chromatography–tandem mass spectrometric (LC–MS/MS) assay for the JAK2 inhibitor CYT387 was developed and validated. Plasma samples were pre-treated using protein precipitation with acetonitrile containing cediranib as internal standard. The extract was directly injected into the chromatographic system after dilution with water. This system consisted of a sub-2 μm particle, trifunctional bonded octadecyl silica column with a gradient using 0.005% (v/v) of formic acid in a mixture of water and methanol. The eluate was transferred into the electrospray interface with positive ionization and the analyte was detected in the selected reaction monitoring mode of a triple quadrupole mass spectrometer. The assay was validated in a 0.25–1000 ng/ml calibration range. Within day precisions were 3.0–13.5%, between day precisions 5.7% and 14.5%. Accuracies were between 96% and 113% for the whole calibration range. The drug was stable under all relevant analytical conditions. Finally, the assay was successfully used to assess drug levels in mice.
Keywords :
JAK2 inhibitor , CYT387 , LC–MS/MS , Human plasma , Mouse plasma
Journal title :
Journal of Chromatography B
Serial Year :
2012
Journal title :
Journal of Chromatography B
Record number :
1469908