Title of article :
Quantitative liquid chromatographic and tandem mass spectrometric determination of vitamin D3 in human serum with derivatization: A comparison of in-tube LLE, 96-well plate LLE and in-tip SPME
Author/Authors :
Xie، نويسنده , , W. and Chavez-Eng، نويسنده , , C.M. and Fang، نويسنده , , W. and Constanzer، نويسنده , , M.L and Matuszewski، نويسنده , , B.K. and Mullett، نويسنده , , W.M. and Pawliszyn، نويسنده , , J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
Sensitive and selective methods based on high performance liquid chromatography (HPLC) with tandem mass spectrometric (MS/MS) detection were developed for the determination of vitamin D3 in human serum. Derivatization of vitamin D3 and its stable isotope labeled internal standard provided highly sensitive quantification and selective detection from endogenous compounds. Samples were prepared using the in-tube liquid–liquid extraction (LLE), 96-well plate LLE, and in-tip solid phase micro-extraction (SPME) in 96-well format. In all methods, the MS/MS detection was performed using Applied Biosystems-Sciex API 3000 tandem mass spectrometers interfaced with a heated nebulizer probe and operated in the positive ionization mode. Both tube and plate LLE methods achieved a lower limit of quantitation (LLOQ) of 0.5 ng/mL when 1.0 and 0.4 mL of human serum was processed, respectively, and were validated in the concentration range of 0.5–25 ng/mL; while for the in-tip SPME method, LLOQ was 5 ng/mL with only 0.1 mL of human serum required. Comparisons were made among three different methods, including precision and accuracy, sample throughput, recovery and matrix effects.
Keywords :
Derivatization , HPLC–MS/MS , vitamin D3 , LLE , SPME , 96-well format , In-tip SPME
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B