Title of article
Quantification of riboflavin in human urine using high performance liquid chromatography–tandem mass spectrometry
Author/Authors
Bishop، نويسنده , , Amanda M. and Fernandez، نويسنده , , Carolina and Whitehead Jr، نويسنده , , Ralph D. and Morales-A، نويسنده , , Pilar and Barr، نويسنده , , Dana Boyd and Wilder، نويسنده , , Lynn C. and Baker، نويسنده , , Samuel E.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
4
From page
1823
To page
1826
Abstract
We developed a selective method to measure riboflavin in human urine. Sample preparation involved solid phase extraction and concentration of the target analyte in urine. The urine concentrate was analyzed using high performance liquid chromatography–tandem mass spectrometry. Riboflavin concentrations were quantified using an isotopically labeled internal standard. The limit of detection was 11 ng/mL, and the linear range was 4.4–20,000 ng/mL. The relative standard deviation at 100, 1000, and 5000 ng/mL was 17%, 17%, and 12%, respectively. The accuracy was 90%. On average, 100 samples, including calibration standards and quality control samples, were prepared per day. Using our method, we measured concentrations of riboflavin in human urine samples that were collected from participants in a study where riboflavin was used as a surrogate chemical to simulate exposure to an environmental toxicant.
Keywords
Riboflavin , Human urine , HPLC–MS/MS , surrogate
Journal title
Journal of Chromatography B
Serial Year
2011
Journal title
Journal of Chromatography B
Record number
1475923
Link To Document