  • Abstract
    We developed a selective method to measure riboflavin in human urine. Sample preparation involved solid phase extraction and concentration of the target analyte in urine. The urine concentrate was analyzed using high performance liquid chromatography–tandem mass spectrometry. Riboflavin concentrations were quantified using an isotopically labeled internal standard. The limit of detection was 11 ng/mL, and the linear range was 4.4–20,000 ng/mL. The relative standard deviation at 100, 1000, and 5000 ng/mL was 17%, 17%, and 12%, respectively. The accuracy was 90%. On average, 100 samples, including calibration standards and quality control samples, were prepared per day. Using our method, we measured concentrations of riboflavin in human urine samples that were collected from participants in a study where riboflavin was used as a surrogate chemical to simulate exposure to an environmental toxicant.
  • Keywords
    Riboflavin , Human urine , HPLC–MS/MS , surrogate