Title of article :
Comparison of different packing methods for capillary electrochromatography columns
Author/Authors :
Roulin، نويسنده , , Stephanie and Dmoch، نويسنده , , Richard and Carney، نويسنده , , Richard and Bartle، نويسنده , , Keith D and Myers، نويسنده , , Peter and Euerby، نويسنده , , Melvin C. and Johnson، نويسنده , , Christopher، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
6
From page :
307
To page :
312
Abstract :
A study was carried out in which 50 μm I.D. fused-silica capillaries were packed with 3 μm octadecylsilane bonded silica, from the same batch, by four methods; liquid slurry and carbon dioxide supercritical carrier, each with and without the use of an ultrasonic probe. A neutral test mixture was analysed by capillary column in reversed-phase mode, and the reproducibility of the electroosmotic flow and of migration time, column efficiency and retention factors, was determined. Initially results suggested that there was no significant difference between properties of columns packed by different methods, and a more thorough statistical evaluation confirmed this; differences observed in the column performance were attributed to random variations between replicate columns, and not between packing methods. However, the variation was least when applying the ultrasonication during liquid slurry.
Journal title :
Journal of Chromatography A
Serial Year :
2000
Journal title :
Journal of Chromatography A
Record number :
1501975
Link To Document :
بازگشت