Title of article
Analysis of heavy metals in atmospheric particulate by ion chromatography
Author/Authors
Bruno، نويسنده , , P. and Caselli، نويسنده , , M. and de Gennaro، نويسنده , , G. and Ielpo، نويسنده , , P. and Traini، نويسنده , , A.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
6
From page
145
To page
150
Abstract
Cu, Ni, Zn, Co, Fe+2, Mn, Cd, Fe+3 and Pb are easily separated and detected in isocratic mode by ion chromatography with post-column derivatization using a bifunctional ion-exchange column and an eluent formed by oxalic acid (28 mM) and sodium nitrate (250 mM). The separation is optimised by using a suggested sample solution containing a given concentration of chloride. Detection limits were 10–15 ppb for all the metals except for cadmium and lead, for which detection limits of 30 and 60 ppb were found, respectively. The method was tested on an atmospheric particulate certified sample. The measured values were in good agreement with certified values. Real samples of atmospheric particulate from industrial and urban sites were analysed and the results are discussed.
Keywords
Heavy metals
Journal title
Journal of Chromatography A
Serial Year
2000
Journal title
Journal of Chromatography A
Record number
1502035
Link To Document