    On-line flow sample stacking in a flow injection analysis–capillary electrophoresis system: 2000-fold enhancement of detection sensitivity for priority phenol pollutants

  • Author/Authors

    Kuban، نويسنده , , Petr and Berg، نويسنده , , Maria and Garc??a، نويسنده , , Carlos and Karlberg، نويسنده , , Bo، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    8
  • From page
    163
  • To page
    170
  • Abstract
    A flow injection analysis–capillary electrophoresis system has been used for on-line flow stacking of 11 US Environmental Protection Agency priority phenol pollutants. Samples containing low concentrations of phenols dissolved in deionised water are continuously delivered to the capillary opening by means of a peristaltic pump. The sample components stack at the boundary between the highly conductive separation electrolyte and the introduced sample. By selecting an appropriate electrolyte and stacking conditions the movement of the electrolyte solution inside the capillary can be reduced, thereby improving the stacking efficiency. The electrolyte used here contained 20 mM phosphate, 8% 2-butanol, and 0.001% hexamethonium bromide at pH 11.95, and the stacking was carried out at 2 kV for 240 s. These conditions allowed up to 2000-fold preconcentration of the selected phenols. No matrix removal was necessary.
  • Keywords
    Phenols , Chlorophenols
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2001