Title of article :
On-line microdialysis coupled with microbore liquid chromatography with ultraviolet detection for continuous monitoring of free cefsulodin in rat blood
Author/Authors :
Tsai، نويسنده , , T.H. and Cheng، نويسنده , , Ian F.C. and Chen، نويسنده , , Y.F. and Chen، نويسنده , , C.F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
5
From page :
83
To page :
87
Abstract :
A microdialysis method followed by a microbore liquid chromatographic ultraviolet detection procedure has been performed for the assay of unbound cefsulodin in rat blood. A microdialysis probe was inserted into the jugular vein for blood sampling. This method involves an on-line design for submitting dialysate into the liquid chromatographic system. The chromatographic conditions consisted of a mobile phase of methanol–100 mM monosodium phosphoric acid (10:90, v/v, pH 5.0) pumped through a microbore reversed-phase column at a flow-rate of 0.05 ml/min. Detection wavelength was set at 265 nm. Microdialysis probes, being laboratory-made, were screened for acceptable in vivo recovery while chromatographic resolution and detection were validated for response linearity as well as intra- and inter-day variabilities. The method was then applied to pharmacokinetics profiling of cefsulodin in the blood following intravenous administration of cefsulodin (20 mg/kg) in rats. Pharmacokinetics were calculated from the corrected data for dialysate concentrations of cefsulodin versus time. Based on pharmacokinetic calculation, cefsulodin best fitted to a two-exponential disposition. This study provided specific pharmacokinetic information for protein-unbound cefsulodin and demonstrated the applicability of this continuous sampling method for pharmacokinetic study.
Keywords :
Cefsulodin , Cephalosporins , antibiotics
Journal title :
Journal of Chromatography A
Serial Year :
2001
Journal title :
Journal of Chromatography A
Record number :
1507705
Link To Document :
بازگشت