Title of article :
Comparison of methods employing gas chromatography retention data to determine vapour pressures at 298 K
Author/Authors :
Koutek، نويسنده , , Bohum??r and Cva?ka، نويسنده , , Josef and Streinz، نويسنده , , Ludv??k and Vrko?ov?، نويسنده , , Pavl??na and Doubsk?، نويسنده , , Jan and ?imonov?، نويسنده , , Hana and Feltl، نويسنده , , Ladislav and Svoboda، نويسنده , , V?clav، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
16
From page :
137
To page :
152
Abstract :
Validity of five models suggested for expressing the relationship between vapour pressures and GC retention times measured on a non-polar capillary column were tested on a common set of compounds [five homologous series of the type H–(CH2)n–Y, where Y denotes Cl, Br, CHO, OCOCH3 and COOCH3, and n varies from 6 to 14]. Standard methods of statistical analysis, as well as vapour pressure values obtained independently from direct vapour pressure measurements were used as validity criteria. For the 40-compound data set examined, the methods provided vapour pressures agreeing within 9.2–24.7% (average absolute percent error) with direct experimental data.
Keywords :
alkanes
Journal title :
Journal of Chromatography A
Serial Year :
2001
Journal title :
Journal of Chromatography A
Record number :
1508109
Link To Document :
بازگشت