• Title of article

    Practical method for evaluation of linearity and effective pathlength of on-capillary photometric detectors in capillary electrophoresis

  • Author/Authors

    Johns، نويسنده , , Cameron and Macka، نويسنده , , Miroslav and Haddad، نويسنده , , Paul R. and King، نويسنده , , Marion and Paull، نويسنده , , Brett، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    5
  • From page
    237
  • To page
    241
  • Abstract
    The optical characteristics of on-capillary photometric detectors for capillary electrophoresis were evaluated and five commercial detectors were compared. Plots of sensitivity (absorbance/concentration) versus absorbance obtained with a suitable testing solution yield both the linear range and the effective path length of the detector. The detector linearity is a crucial parameter when using absorbing electrolytes, such as for indirect photometric detection, and especially for highly absorbing electrolyte probe ions. The upper limits of the linear ranges (determined as 5% decline in sensitivity) for five commercial detectors ranged from 0.175 to 1.2 AU. The effective pathlength reflects the quality of the optical design of the detector and is equal to the capillary internal diameter only for a light beam passing exactly through the capillary centre, but becomes progressively shorter for imperfect optical designs. The determined effective pathlength for the five investigated detectors ranged from 49.7 to 64.6 μm for a 75 μm I.D. capillary.
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography A
  • Record number