• Title of article

    The ultimate band compression factor in gradient elution chromatography

  • Author/Authors

    Gritti، نويسنده , , Fabrice and Guiochon، نويسنده , , Georges، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    13
  • From page
    79
  • To page
    91
  • Abstract
    The equations predicting the ultimate time band compression factor, G = ( t R − t F ) / t p in linear gradient elution chromatography, for an infinitely narrow injection (injection time t p → 0 ) were derived for an ideal-model column (dispersionless chromatography, H = 0 ) assuming the Linear Solvent Strength Model for the retention behavior of the analyte. Numerical solutions can readily be obtained when the LSSM model does not apply. The results can be generalized to any retained organic modifier ( k A ′ ) in the mobile phase. The stronger the retention of the organic modifier, the more effective the band compression. Dispersion in real chromatographic column ( H ≠ 0 ) affects the limits that can be reached in linear gradients but poorly in step gradients. Examples based on a conventional HETP of about 12 μ m using a 5 μ m particle packed column reveal that the best time compression factor that could be expected is twice the one predicted with an ideal column.
  • Keywords
    caffeine , Column efficiency , General isotherm modeling , Peak compression factor , RP-HPLC stationary phases , Separation , Acetonitrile–water mobile phases , Analytical chromatography , Band compression , Gradient elution chromatography
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography A
  • Record number