Title of article
Improved method for direct high-performance liquid chromatography assay of angiotensin-converting enzyme-catalyzed reactions
Author/Authors
Wu، نويسنده , , Jianping and Aluko، نويسنده , , Rotimi E and Muir، نويسنده , , Alister D، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
6
From page
125
To page
130
Abstract
A rapid and sensitive assay was developed for determination of the activity of angiotensin-converting enzyme (ACE) in the presence of inhibitory peptides present in soybean protein hydrolysates. The method utilizes reversed-phase high-performance liquid chromatography (HPLC) to separate and quantify hippuryl-histidyl-leucine (HHL) and hippuric acid (HA). HHL and HA were separated on a Symmetry C18 column by gradient elution that used mixtures of trifluoroacetic acid (TFA)–acetonitrile and TFA–water as solvents. Analytical time and baseline separation of HA from HHL were improved over previous HPLC methods. In comparison to the standard spectrophotometric method, the new HPLC method obviates the need for ethyl acetate extraction of HA but requires direct injection of the ACE reaction mixture onto the HPLC column.
Keywords
Proteins , Enzyme inhibitors , Hippuric Acid , Peptides , angiotensin-converting enzyme , enzymes
Journal title
Journal of Chromatography A
Serial Year
2002
Journal title
Journal of Chromatography A
Record number
1510294
Link To Document