Title of article :
Fabrication and evaluation of single- and dual-channel (Π-design) microchip electrophoresis with electrochemical detection
Author/Authors :
Mario and Pozo Ayuso، نويسنده , , Diego F. and Castaٌo-ءlvarez، نويسنده , , Mario and Fernلndez-la-Villa، نويسنده , , Ana and Garcيa-Granda، نويسنده , , Miguel and Fernلndez-Abedul، نويسنده , , M. Teresa and Costa-Garcيa، نويسنده , , Agustيn and Rodrيguez-Garcيa، نويسنده , , Jose، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
A new dual-channel microchip capillary electrophoresis (MCE) has been developed on glass substrates for the first time with electrochemical detection. Dual-channel (called Π-design) as well as single-channel microchips have been fabricated on soda-lime glass using photolithography, wet etching and thermal bonding. Moreover, a laser writing system has been applied for the fabrication of photomasks with the different microchip designs (single- and dual-channel configurations). The microfabricated channels have been characterized by optical, confocal and scanning electron microscopy. The resulting single- and dual-channel microchips have been evaluated using an end-channel amperometric detector based on one (single-channel) or two (dual-channel) 100-μm gold wires aligned at the outlet of the separation channel. Parameters affecting the separation of several phenolic compounds (dopamine, p-aminophenol and hydroquinone) have been studied in the glass microchips. Thus, the influence of separation voltage, detection potential and background electrolyte has been examined in the single-channel microchip. Different total length microchannel has been compared. Furthermore, the possibility of carrying out two simultaneous measurements has been demonstrated in the new dual-channel microchip electrophoresis. The injection format has been checked and resulted to be critical, in such a way that a special and new form is employed for obtaining simultaneous signals at both channels. Analytical characteristics, such as sensitivity and reproducibility have been evaluated and resulted very adequate.
Keywords :
Glass microchip electrophoresis , Photolithography , amperometric detection , Dual-channel , End-channel gold wire , Wet etching
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A