Title of article :
A two-dimensional suspension array system by coupling field flow fractionation to flow cytometry
Author/Authors :
Li، نويسنده , , Jishan and Zhong، نويسنده , , Wenwan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
7
From page :
143
To page :
149
Abstract :
Flow field flow fractionation (Fl-FFF) was coupled to flow cytometry to improve the performance of suspension arrays. Size-based separation of the protein-conjugated microspheres by Fl-FFF was performed and the results demonstrated that, the separation could tolerate a wide range of carrier fluid conditions (pH values, salt concentrations, and buffer compositions) favorable for immunoassays. The immuno-complex remained intact during Fl-FFF, as revealed by fluorescence measurements before and after the Fl-FFF separation, and SDS-PAGE of the eluted proteins. The sample throughput of the suspension array can be increased several folds by using particles of different sizes and separating them with Fl-FFF before flow cytometric measurement. Moreover, the gel result hinted that the continuous wash inside the Fl-FFF system may lower the assay background, another possible advantage of the two-dimensional suspension array system.
Keywords :
Flow field flow fractionation , Size separation , Suspension array , Array throughput , Immunoassay
Journal title :
Journal of Chromatography A
Serial Year :
2008
Journal title :
Journal of Chromatography A
Record number :
1510478
Link To Document :
بازگشت