Title of article
Development of a supercritical fluid chromatography high-resolution separation method suitable for pharmaceuticals using cyanopropyl silica
Author/Authors
Brunelli، نويسنده , , Claudio and Zhao، نويسنده , , Yining and Brown، نويسنده , , Melissa-Hanna and Sandra، نويسنده , , Pat، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
10
From page
263
To page
272
Abstract
A method has been developed for the analysis of a broad spectrum of pharmaceuticals using packed column supercritical fluid chromatography (pSFC) on a cyanopropyl silicagel stationary phase. Five 25 cm × 4.6 mm I.D., 5.0 μm columns were coupled to generate ca. 100 000 plates. The selectivity was tuned by varying the nature and concentration of various modifiers and additives in the carbon dioxide mobile phase. It was noted that pressure influences both efficiency and selectivity of the chromatographic process. Final method conditions are: outlet pressure 100 bar, flow 2.0 mL/min, temperature 40 °C, organic modifier program from 5% (1 min) to 40% at 2.0%/min, organic modifier composition methanol:acetonitrile in a ratio of 3:1 (variable according to sample composition) with peak symmetry additives trifluoroacetic acid and diisopropylamine both at levels of 0.5%.
Keywords
Resolution optimization , Packed column supercritical fluid chromatography , Pharmaceutical Analysis , high efficiency , Cyanopropyl silicagel
Journal title
Journal of Chromatography A
Serial Year
2008
Journal title
Journal of Chromatography A
Record number
1510566
Link To Document