• Title of article

    Determination of pipecuronium bromide and its impurities in pharmaceutical preparation by high-performance liquid chromatography with coulometric electrode array detection

  • Author/Authors

    B?a?ewicz، نويسنده , , Agata and Fija?ek، نويسنده , , Zbigniew and Samsel، نويسنده , , Karolina، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    191
  • To page
    195
  • Abstract
    A sensitive and selective HPLC method with coulometric electrode array detection for the determination of pipecuronium bromide and its four impurities has been developed. The coulometric electrode array detection at increasing potentials from +300 to +900 mV of the porous graphite electrode versus the palladium reference electrode was used. The limit of detection and quantitation for pipecuronium bromide was 8 and 25 ng ml−1, respectively. This elaborate method for the simultaneous analysis of pipecuronium bromide and its impurities proved to be fast, precise, accurate, sensitive, and could be applied to analysis in substances and in pharmaceutical preparations.
  • Keywords
    Electrochemical detection , Coulometric electrode array detection , Impurities , Pipecuronium bromide
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography A
  • Record number