• Title of article

    Kinetic plot equations for evaluating the real performance of the combined use of high temperature and ultra-high pressure in liquid chromatography: Application to commercial instruments and 2.1 and 1 mm I.D. columns

  • Author/Authors

    Heinisch، نويسنده , , Sabine and Desmet، نويسنده , , Gert and Clicq، نويسنده , , David and Rocca، نويسنده , , Jean-Louis، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    13
  • From page
    124
  • To page
    136
  • Abstract
    The use of ultra-high pressure liquid chromatography (UHPLC) with pressures up to 1000 bar and columns packed with sub-2-μm particles combined with high-temperature mobile phase conditions (up to 90 °C) is assessed according to the current available instrumentation via constrained kinetic plot equations. It is shown that the gain in separation speed, theoretically expected from high-temperature UHPLC (HT-UHPLC), is significantly reduced when taking into account the existing instrumental constraints (extra-column band broadening, flow-rate and column length limitations). This study also shows that significant improvements could be expected on the current commercial instruments by increasing the flow-rate limit and/or using packing columns with particle size in the range 2.5–3.5 μm instead of the current sub-2 μm. These particles should obviously withstand very high pressure.
  • Keywords
    high temperature , Ultra-high pressure , Liquid chromatography , Kinetic plot method
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography A
  • Record number

    1511142