Title of article
Use of absorbent materials in on-line coupled reversed-phase liquid chromatography–gas chromatography via the through oven transfer adsorption desorption interface
Author/Authors
Flores، نويسنده , , Gema and Dيaz-Plaza، نويسنده , , Eva Marيa and Cortés، نويسنده , , Jose Manuel and Villén، نويسنده , , Jesْs and Herraiz، نويسنده , , Marta، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
5
From page
99
To page
103
Abstract
The use of absorbents as retaining materials in the through oven transfer adsorption desorption interface (TOTAD) of an on-line coupled reversed-phase liquid chromatography–gas chromatography system (RPLC–GC) is proposed for the first time. A comparative study of an adsorbent (Tenax TA) and two absorbents, namely polydimethylsiloxane and poly(50% phenyl/50% methylsiloxane) is performed to establish the best experimental conditions for the automated and simultaneous determination of 15 organophosphorus and organochlorine pesticide residues in olive oil. The proposed method provides satisfactory repeatability (RSDs lower, in general, than 8.5%) and sensitivity (limits of detection ranging from 0.6 to 81.9 μg/L) for the investigated compounds.
Keywords
Absorbents , adsorbents , Pesticide residue analysis , TOTAD interface , On-line coupling RPLC–GC
Journal title
Journal of Chromatography A
Serial Year
2008
Journal title
Journal of Chromatography A
Record number
1511382
Link To Document