Title of article
Simple and precise detection of lipid compounds present within liposomal formulations using a charged aerosol detector
Author/Authors
Schِnherr، نويسنده , , Carina and Touchene، نويسنده , , Sounia and Wilser، نويسنده , , Gaston and Peschka-Süss، نويسنده , , Regine and Francese، نويسنده , , Giancarlo، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
6
From page
781
To page
786
Abstract
In recent decades the use of liposomal preparations as drug delivery systems has become very attractive in pharmaceutical development. Therefore, thorough characterization and quantification of the lipids which form liposomes is wished from both investigators and regulatory authorities when the application in humans is being considered. In this study a new HPLC method for the detection of lipids in liposomal formulations was established using corona charged aerosol detection (CAD) which has the advantage to be independent of the chemical properties of the analytes. The superiority of this method over UV detection was demonstrated. Compared to UV detection no absorption effects of the organic solvent in the mobile phase interfering with the lipid signals were observed with CAD. CAD showed good linearity (R2 > 0.990) for all liposomal compounds. The acceptance criteria for precision including repeatability were met. The average recovery for each of the excipients of the liposomal formulation was in the range of 90.0–110%.
Keywords
Corona aerosol detector (CAD) , HPLC , Phosphatidylcholine , Lipid detection , Liposomes
Journal title
Journal of Chromatography A
Serial Year
2009
Journal title
Journal of Chromatography A
Record number
1511588
Link To Document